ultrasound protocol to assess tips
Transjugular Intrahepatic Portosystemic Shunt (TIPS)
Assess:
- Portal vein with real-time, PW and colour Doppler for:
- Size of vessel
- Presence of thrombus
- Direction of flow
- Maximum velocity (Vmax) of flow prior to stent
- TIPS stent with real-time, PW and colour Doppler for:
- Vmax in stent (careful assessment of velocities throughout the stent is required
- Vmax in portal 1/3 of stent
- Evidence of narrowing of lumen (reduced lumen in real-time or colour, turbulent colour pattern)
- Hepatic vein with real-time, PW and colour Doppler for:
- Increase velocity beyond stent
- Presence of thrombus or narrowing of vein beyond stent
- IVC with real-time, PW and colour Doppler for:
- Normal flow patterns
- Presence of thrombus
Note: visualization of the hepatic vein and IVC may be limited or impossible in a significant number of patients
Standard Documentation:
Essential:
- Portal vein prior to stent
- Long axis with lumenal diameter / clip store with colour Doppler
- Spectral trace with Vmax
TIPS Stent - Long axis views low, mid and high / clip store with colour Doppler
- Spectral trace in portal 1/3 of stent with Vmax measured
- Spectral trace of the highest velocity found in the stent with Vmax measured
Acquire if possible:
- Hepatic vein
- Long axis / clip store with colour Doppler
- Spectral trace with Vmax
- IVC
- Long axis / clip store with colour Doppler
- Spectral trace with Vmax
|
Go back to ultrasound protocols homepage