Nuchal TranslucencyThis is a featured page

Ultrasound scanning guidelines

Nuchal Translucency


Performed on patients as part of a 1st trimester screening programme
Scan transabdominally initially, transvaginal scanning can be considered if there is difficulty in obtaining correct fetal position

Fetus must be between 45mm and 82mm Crown Rump Length in size

Imaging Criteria:

Minimum criteria for selecting a NT measurement as adequate include:

  • Fetus must be supine
  • Measurement must be in a true sagittal section, landmarks are fetal face and heart
  • Magnification so that head and chest take up 75% of the image
  • Head in neutral position (mild flexion is acceptable but the chin must not be on the chest wall)
Ideal but not essential

  • A measurement with the fetus lifted off the amnion
  • Fetal head in true neutral position
Measurement:

  • Measure the widest point perpendicular to the skinline
  • If the cord is around the neck the measurement can be falsely increased, in this case take the average of the measurement above and below the cord
  • Record the largest measurement from the best image
Documentation:

  • Long and trans image of the uterus
  • Crown rump length
  • Fetal heart
  • Nuchal translucency measurement x2 meeting the above criteria

....Go back to ultrasound protocols homepage


AndyC
AndyC
Latest page update: made by AndyC , Feb 25 2009, 5:05 AM EST (about this update About This Update AndyC Edited by AndyC

195 words added

view changes

- complete history)
More Info: links to this page
There are no threads for this page.  Be the first to start a new thread.

Related Content

  (what's this?Related ContentThanks to keyword tags, links to related pages and threads are added to the bottom of your pages. Up to 15 links are shown, determined by matching tags and by how recently the content was updated; keeping the most current at the top. Share your feedback on Wetpaint Central.)