Mobile CXR scatter radiationThis is a featured page

Scatter Radiation Exposure From Mobile Chest X-Rays

The calculated scattered radiation dose to persons in the proximitiy of patients receiving a chest x-ray from mobile xray units assuming the following typical operating conditions

  • 85 kVp, 2 mAs, 3mm Al filtration
  • 180 cm FFD
  • 35cm x 43cm CR image receptor
  • Proximity of the nearest person on either side of patient is around 1 metre or more,and the primary beam is completely intercepted by the receptor and not directed towards other patients or staff.

Under these conditions the entrance exposure (air kerma) to the patient is calculated to be 52 micrograys, and the scattered radiation at one metre to either side around 0.2 micrograys. To place this in perspective it should be borne in mind that the dose from natural background radiation (to which everyone is exposed) is of the order of 2400 micrograys per year; the scattered radiation from a mobile CXR at a distance of 1 metre is therefore less than that received from one hour of natural background radiation. This is an extremely small radiation exposure and the resultant risk to staff and patients is negligible.

These calculations were performed by the
Chief Medical Scientist at Flinders Medical Centre
Radiation Safety Officer
Division of Medical Imaging


Further reading



...Go back to General Radiography Homepage
...Go back to wikiRadiography mainpage



Links



XrayHomer
XrayHomer
Latest page update: made by XrayHomer , Oct 18 2010, 3:38 PM EDT (about this update About This Update XrayHomer Edited by XrayHomer


view changes

- complete history)
More Info: links to this page
There are no threads for this page.  Be the first to start a new thread.

Related Content

  (what's this?Related ContentThanks to keyword tags, links to related pages and threads are added to the bottom of your pages. Up to 15 links are shown, determined by matching tags and by how recently the content was updated; keeping the most current at the top. Share your feedback on WikiFoundry Central.)