Exposure - RibsThis is a featured page

This page contains radiographic exposure values

Adult Ribs

Region View kVp mAs Cassette Grid Distance
RibsAP Oblique6616RegularYes100 - 115cm (40 inches)

AP (Posterior Upper Ribs 1-8)6616RegularYes100 - 115cm (40 inches)

AP (Posterior Lower Ribs 9-12)7530RegularYes100 - 115cm (40 inches)

PA (Anterior Ribs6616RegularYes100 - 115cm (40 inches)
ChestPA1004RegularYes180cm (71 inches)

Paediatric Ribs

Region View kVp mAs Cassette Grid Distance
RibsAP Oblique6012RegularYes100 - 115cm (40 inches)

AP (Posterior Upper Ribs 1-8)6012RegularYes100 - 115cm (40 inches)

AP (Posterior Lower Ribs 9-12)7020RegularYes100 - 115cm (40 inches)

PA (Anterior Ribs6012RegularYes100 - 115cm (40 inches)
ChestPA902RegularYes180cm (71 inches)


These are average exposures using a Carestream CR system, exposures may vary dramatically between different film/screen combinations, CR or DR systems


....Go Back to Exposure Values homepage



AndyC
AndyC
Latest page update: made by AndyC , Jun 7 2011, 1:51 AM EDT (about this update About This Update AndyC Edited by AndyC

6 words added
4 words deleted

view changes

- complete history)
Keyword tags: exposure ribs
More Info: links to this page
There are no threads for this page.  Be the first to start a new thread.

Related Content

  (what's this?Related ContentThanks to keyword tags, links to related pages and threads are added to the bottom of your pages. Up to 15 links are shown, determined by matching tags and by how recently the content was updated; keeping the most current at the top. Share your feedback on Wetpaint Central.)